Comparison between the QRMA Measurement with the Anamnesis and the Capillary Blood Glucose Test
Downloads
ABSTRACT
Introduction: The examination of patients with diabetes mellitus (DM) can be done by reviewing their complaints and through a capillary blood glucose level test to determine the value of their Random Blood Glucose Level. QRMA (Quantum Resonance Magnetic Analyzer) is claimed to be able to check the patient's bodily condition (including blood glucose) with an accuracy of 85%. The purpose of this study was to verify the validity of the QRMA tool and its accuracy by comparing the results of the anamnesis and the examination conducted using the capillary blood glucose test method.
Methods: The research method used was a cross-sectional design. The total sample consisted of 44 respondents in the working area of the Community Health Centers in Yogyakarta with the risk factor being blood sugar level instability. The sampling technique used was purposive sampling. The main variable in this study was the value of the blood sugar level measured based on the coefficient value of the QRMA tool and the value of Random Blood Glucose obtained through the capillary blood glucose test.
Results: The blood glucose value was not correlated significantly with the coefficient value of QRMA. The value of blood glucose when examined alongside the result of the respondent's anamnesis showed there to be a significant difference. The value of the QRMA coefficient when examined against the results from the history of the respondents showed no significant difference. Linear regression showed that the variables of height, body weight, and IMT had a correlation with the QRMA coefficient value.
Conclusion: TheQRMA tool was not able to provide a picture of the actual condition of the blood glucose level of the respondents when compared with the results of the anamnesis and the blood glucose value from the capillary blood glucose test. Non-invasive health measurement devices such as QRMA are not used by nurses as a standard for determining the health status of DM patients.
American Diabetes Association. (2014). Diagnosis And Classification Of Diabetes Mellitus. Diabetes Care, 37((Supplement 1) January), S81–S90. Https://Doi.Org/10.2337/Dc14-S081
Boyd, R., Leigh, B., & Stuart, P. (2005). Capillary Versus Venous Bedside Blood Glucose Estimations. Emergency Medicine Journal, 22(3), 177–179. Https://Doi.Org/10.1136/Emj.2003.011619
Chirwa, L. C., Hammond, P. A., Member, S., Roy, S., & Cumming, D. R. S. (2003). Electromagnetic Radiation From Ingested Sources In The Human Intestine Between 150 Mhz And 1 . 2 Ghz. IEEE Transactions On Biomedical Engineering, 50(4), 484–492.
Critchell, C. D., Savarese, V., Callahan, A., Aboud, C., Jabbour, S., & Marik, P. (2007). Accuracy Of Bedside Capillary Blood Glucose Measurements In Critically Ill Patients. Intensive Care Medicine, 33(12), 2079–2084. Https://Doi.Org/10.1007/S00134-007-0835-4
Handelsman, Y., Bloomgarden, Z. T., Grunberger, G., Umpierrez, G., Zimmerman, R. S., Bailey, T. S., ... Zangeneh, F. (2015). American Association Of Clinical Endocrinologists And American College Of Endocrinology – Clinical Practice Guidelines For Developing A Diabetes Mellitus Comprehensive Care Plan – 2015. Endocrine Practice, 21((Suppl 1) April), 1–87.
Hill, N. R., Oliver, N. S., Choudhary, P., Levy, J. C., Hindmarsh, P., & Matthews, D. R. (2011). Normal Reference Range For Mean Tissue Glucose And Glycemic Variability Derived From Continuous Glucose Monitoring For Subjects Without Diabetes In Different Ethnic Groups. Diabetes Technology & Therapeutics, 13(9), 921-928.
Holmer, H. K., Ogden, L. A., Burda, B. U., & Norris, S. L. (2013). Quality Of Clinical Practice Guidelines For Glycemic Control In Type 2 Diabetes Mellitus. Plos One, 8(4), 1–6. Https://Doi.Org/10.1371/Journal.Pone.0058625
Kemenkes RI. (2018). Hasil utama RISKESDAS 2018. Jakarta: Kementrian Kesehatan Badan Penelitian dan Pengembangan Kesehatan.
Kotwal, N., & Pandit, A. (2012). Variability Of Capillary Blood Glucose Monitoring Measured On Home Glucose Monitoring Devices. Indian J Endocrinol Metab, 16(Supplement 2), 248–251.
Kulkarni, A., Saxena, M., Price, G., O'Leary, M. J., Jacques, T., & Myburgh, J. A. (2005). Analysis Of Blood Glucose Measurements Using Capillary And Arterial Blood Samples In Intensive Care Patients. Intensive Care Medicine, 31(1), 142–145. Https://Doi.Org/10.1007/S00134-004-2500-5
Kurniawan, B., & Wahyuni, I. (2008). Hubungan Radiasi Gelombang Elektromagnetik Dan Faktor Lain Dengan Keluhan Subyektif Pada Tenaga Kerja Industri Eletronik GE Di Yogyakarta. Jurnal Promosi Kesehatan Indonesia, 3(2), 127–133.
Larin, K. V., Eledrisi, M. S., Motamedi, M., & Esenaliev, R. O. (2002). Noninvasive Blood Glucose Monitoring With Optical Coherence Tomography. Diabetes Care, 25(12), 2263–2267.
Liu, D., Moberg, E., Kollind, M., Lins, P. E., Adamson, U., & Macdonald, I. A. (1992). Arterial, Arterialized Venous, Venous And Capillary Blood Glucose Measurements In Normal Man During Hyperinsulinaemic Euglycaemia And Hypoglycaemia. Diabetologia, 35(3), 287–290.
Management Of Diabetes Mellitus Update Working Group. (2010). VA/Dod Clinical Practice Guideline For The Management Of Diabetes Mellitus (Version 4.). Washington.
Mazze, R. S., Ph, D., Pasmantier, R., Lucido, D., Ph, D., Murphy, J., ... Ph, D. (1994). Reliability Of Blood Glucose Monitoring By Patients With Diabetes Mellitus, 77(August), 211–217.
Montagnana, M., Caputo, M., Giavarina, D., & Lippi, G. (2009). Overview On Self-Monitoring Of Blood Glucose. Clinica Chimica Acta, 402(1-2), 7-13.
Muktabhant, B., Sanchaisuriya, P., Sarakarn, P., Tawityanon, W., Trakulwong, M., Worawat, S., & Schelp, F. P. (2012). Use Of Glucometer And Fasting Blood Glucose As Screening Tools For Diabetes Mellitus Type 2 And Glycated Haemoglobin As Clinical Reference In Rural Community Primary Care Settings Of A Middle Income Country. BMC Public Health, 12(1), 349.
Niwinski, N. (2009). Capillary Blood Collection: Best Practices. Labnotes, 20(1), 1–8.
Rajaratnam, H. N., & Pathmanathan, S. (2011). How Reliable Are Capillary Blood Glucose Measurements ? Sri Lanka Journal Of Diabetes Endocrinology And Metabolism, 1(1), 22–24.
Rantung, J., Yetti, K., & Herawati, T. (2015). Hubungan Self-Care Dengan Kualitas Hidup Pasien Diabetes Melitus (Dm) Di Persatuan Diabetes Indonesia (Persadia) Cabang Cimahi. Jurnal Skolastik Keperawatan, 1(1), 38–51.
Redmon, B., Caccamo, D., Flavin, P., Michels, R., O'Connor, P., Roberts, J., ... Sperl-Hillen, J. (2014). Diagnosis And Management Of Type 2 Diabetes Mellitus In Adults. Institute For Clinical Systems Improvement, 16(7), 1–85.
Schifman, R. B., Nguyen, T. T., & Page, S. T. (2014). Reliability Of Point-Of-Care Capillary Blood Glucose Measurements In The Critical Value Range. Archives Of Pathology And Laboratory Medicine, 138(7), 962-966.
Suyatno, F. (2008). Aplikasi Radiasi Sinar-X Di Bidang Kedokteran Untuk Menunjang Kesehatan Masyarakat. In Seminar Nasional IV SDM Teknologi Nuklir (Pp. 25–26). Yogyakarta.
Swamardika, I. B. A. (2009). Pengaruh Radiasi Gelombang Elektromagnetik Terhadap Kesehatan Manusia. Majalah Ilmiah Teknologi Elektro, 8(1).
Tirimacco, R., Tideman, P. A., Dunbar, J., Simpson, P. A., Philpot, B., Laatikainen, T., & Janus, E. (2010). International Journal Of Diabetes Mellitus Should Capillary Blood Glucose Measurements Be Used In Population Surveys ? International Journal Of Diabetes Mellitus, 2(1), 24–27. Https://Doi.Org/10.1016/J.Ijdm.2009.12.002
Tjokroprawiro, A. (2007). Buku Ajar Ilmu Penyakit Dalam. Surabaya: Airlangga University Press.
Tonyushkina, K., & Nichols, J. H. (2009). Glucose Meters: A Review Of Technical Challenges To Obtaining Accurate Results, 3(4).
Victorya, R. M. (2015). Effects Of Handphone ' S Electromagnetic Wave Exposure On Seminiferous Tubules. Jurnal Majority, 4(3).
WHO. (2010). WHO guidelines on drawing blood : best practices in phlebotomy. Genewa: WHO Document Production Services.
Wild, S. H., Roglic, G., Green, A., Sicree, R., & King, H. (2004). Global prevalence of diabetes: estimates for the year 2000 and projections for 2030: response to Rathman and Giani. Diabetes Care, 27(10), 2569.
Authors who publish with Jurnal Ners agree to the following terms:
- Authors transfer the Copyright and grant Jurnal Ners the right of first publication with the work simultaneously licensed under a Creative Commons Attribution 4.0 International License that allows others to remix, adapt and build upon the work with an acknowledgment of the work's authorship and of the initial publication in Jurnal Ners.
- Authors are permitted to copy and redistribute the journal's published version of the work (e.g., post it to an institutional repository or publish it in a book), with an acknowledgment of its initial publication in Jurnal Ners.
Jurnal Ners requires a formal written declaration and transfer of copyright from the author(s) for each article published. We, therefore, ask you to complete and return this form, retaining a copy for your own records. Your cooperation is essential and appreciated. Any delay will result in a delay in publication. The form can be downloaded HERE.