Optimalization Image Of Turbo Spin Echo (TSE) With Pre Saturation And Gradient Moment Nulling (GMN) To Reduce Flow Artifact On MRI Cervical
Downloads
ABSTRACT
The research of Optimization Image of Turbo Spin Echo (TSE) with Pre saturation and Gradient Moment Nulling (GMN) to reduce flow artifact on MRI Cervical has been done. The purpose of this research is to know the effect of pre saturation and gradient moment nulling (GMN) on cervical MRI and determine optimal image to reduce flow artifact. This research used four treatment variations that were without pre saturation and without flow compensation (GMN), with pre saturation, with flow compensation (GMN), and with both pre saturation and flow compensation (GMN) on sequence T2-weighting TSE sagital on cervical MRI. The quantitative analysis done by using Region of Interest (ROI) on MRI image then analyzed signal to Noise Ratio (SNR) and Contrast to Noise Ratio (CNR). The best effect and image quality obtained by pre saturation and flow compensation (GMN) treatment with SNR value on tissue was 328,7 at vertebra cervical, 278,6 at spinal cord, 366,6 at discus, 596,3 at CSF. While CNR tissue value was 78,6 at vertebra cervical, 257,6 at spinal cord, 274,8 at discus, and
336,3 at CSF followed by the decrease of flow artifact in the amount of 160,4. Conclusion shows that the image with both pre saturation and flow compensation (GMN) treatment on T2 TSE sagital can reduce flow artifact signal on the spinal cord tissue.
Keywords : Pre saturation, flow compensation, gradient moment nulling (GMN), MRI Cervical.
DAFTAR PUSTAKA
Blink, E. J. 2009. Mri : Physics, Netherlands 0-75.
Brown, M. A., Semelka, R. C. 2003. Mri: Basic Principles And Applications, Third Edition. John Wiley And Sons, Inc : New Jersey.
Dance, D. R., Maidment, A. D., Christofides, S., McLean, I. D.,Ng, K. 2014.
Diagnostic Radiology Physics A Handbook for Teachers and Students. International Atomic Energy Agency (IAEA): Vienna.
Deviyanti, M. E., Sugiyanto, Indrati, R. 2011. Studi Perbandingan Artefak Antara Mri Lumbal T1 Spin Echo Dengan Pre Saturasi Dan T1 Spin Echo Tanpa Pre Saturasi. E-Journal Politeknik Kesehatan Semarang: Semarang.
Felmlee, Joel. P., Ehman, R.L,. 1986. Spatial Presaturation: A Method for Suppressing Flow Artifact and Improving Depiction of Vascular Anatomy in MR Imaging. Departement of Radiology: Rochester.
Gray, Henry. 1960. Anatomy of the Human Body. Columbia.
Gaa, J., Rummeny, E. J., & Seemann, M. D. 2004. Whole-Body Imaging With
Pet/Mri. European Journal Of Medical Research: Germany.
Hinks, R. Scott., Constable, R. Todd. 1994. Gredient Moment Nulling in Fast Spin
Echo. Williams & Wilkins : Canada.
Jahng, Geon-Ho., Pickup, Stephen. 2000. Gradient Optimized Gradient-Echo
Gradient Moment Nulling Sequences for Flow Compensation of Brain
Images. Departement of Diagnostic Imaging and Therapeustics BSAC
Kraugusteeliana, Napitupulu, P. 2009. Penerapan Citra Medik Pada Visualisasi Pencitraan Diagnostik Tumor Otak Secara Sagittal Mengunakan Magnetic Resonance Imaging ( MRI ). Politeknik Negeri Semarang : Semarang.
Muzammil, Akhmad. 2015. MRI, Second Edition. Surabaya.
Rochmayanti, D., Widodo, T. S., Soesanti, I. 2013. Analisis Perubahan Parameter Number Of Signals Averaged ( NSA ) Terhadap Peningkatan SNR Dan Waktu Pencitraan Pada MRI. Universitas Gajah Mada : Yogyakarta.
Siemens, 2012. Global Siemens Healthcare :Magnetom Essenza. Copyrigth © Siemens AG.
Simanjuntak, J. N., Nur, M., Hidayanto, E. 2014. Studi Analisis Echo Train Length dalam K- Space Serta Pengaruhnya Terhadap Kualitas Citra Pembobotan T2 FSE. Berkala Fisika Universitas Gajah Mada : Yogyakarta.
Soesanti, I., Susanto, A., Sri, T., Maesadji, W. 2011. Ekstraksi Ciri dan Identifikasi Citra Otak MRI Berbasis Eigenbrain Image. Universitas Gajah Mada : Yogyakarta.
Soetikno, Ristaniah, D. 20. Imejing Molekuler Menggunakan MRI: Cara Baru untuk Diagnosis Tumor Otak Glikoma. Departemen Radiologi FK UNPAD: Bandung.
Somasundaram, K., Kalavathi, P. 2012. Analysis of Imaging Artifacts in MR Brain Images. Oriental Journal Of Computer Science & Technology : India.
Westbrook, Catherine., Kaut, Carolyn. 1998. MRI in Practice, Second Edition.
Blackwell Science Ltd: London.
Westbrook, Catherine. 2014. Handbook of MRI Technique Fourth Edition. John
Wiley & Sons
Authors who publish with this journal agree to the following terms:
- The copyright of this journal belongs to the Editorial Board and Journal Manager with the author's knowledge, while the moral right of the publication belong to the author.
- The formal legal aspect of journal publication accessibility refers to the Creative Commons Attribution-Share Alike (CC BY-SA).
- Every publication (print/electronic) is open access for educational, research, and library purposes. In addition to the objectives mentioned above, the editorial board is not responsible for copyright infringement